METER CALIBRATION
METER CALIBRATION PEOCEDURE
•
••
•
Connect transceiver to TRANSMITTER IN connector.
•
••
•
Connect external 50
Ω
load to COAX 2 connector.
•
••
•
Set the OUTPUT SELECT switch to the COAX 2 DIRECT position.
•
••
•
Set the RANGE button to 300 W and the PEAK/AVG button to AVG.
•
••
•
Apply 100W of RF at 14.0 MHz.
•
••
•
Adjust R1 (see Figure 1) so that 100W of FORWARD power is read on the
meter.
•
••
•
Reduce the RF power to 10W and set the RANGE to 30W
•
••
•
Adjust R2 so that 10W of FORWARD power is read on the meter.
•
••
•
Reverse the transceiver and load connectors on the rear panel
•
••
•
Set the RANGE to 300W.
•
••
•
Apply 10W of RF power and adjust R4 to read 10W of REFLECTED power.
•
••
•
Set the RANGE to 30W.
•
••
•
Apply 2W of RF power and adjust R3 to read 2W of REFLECTED power.